THMS600-PS Specifications


Specifications at 14bar

  • Temperature Range -100°C to 500 °C (14Bar), -196°C to 600°C (atmospheric)

  • Sample area of 24.5mm Ø

  • 16 mm X,Y sample manipulation

  • Up to 100 temperature profiles can be programmed

  • Option between touchscreen LinkPad or PC

  • Stage can be pressurised up to 14Bar

  • Can be used with transmitted or reflected light

  • Mounts directly to microscope table or substage

  • Stage body size - 135 x 92 x 25.1 mm

  • 100 Ohm platinum sensor

  • Temperature stability < 0.1°C

  • Direct injection of coolant into block

  • Highly conductive metal for improved heat transfer

  • Maximum heating of 150°C/min

  • Objective lens minimum working distance 8.7mm

  • Condenser lens minimum working distance 14.5mm

If you are using transmitted light please go to the CONDENSER EXTENSION LENS section for condenser info.