THMS600-PS Specifications


Specifications at 14bar

  • Temperature Range -120°C to 500 °C
  • Sample area of 22mm diameter
  • 16 mm X,Y sample manipulation
  • 32 ramp temperature profile programming
  • No PC required
  • Pressure tested to 14bar
  • Can be used with transmitted or reflected light
  • Mounts directly to microscope table or substage
  • Stage body size - 160 x 80 x 24 mm
  • 100 ohm platinum sensor 1/10th din standard class A
  • Temperature stability < 0.1°C
  • Direct injection of coolant into block
  • Highly conductive metal for improved heat transfer
  • Maximum heating of 150°C/min
  • Response time of <1 second at 5°C/min at 50°C
  • Objective lens minimum working distance 6.6mm
  • Condenser lens minimum working distance 14.5mm

If you are using a light microscope, please go to the STAGE CLAMPS section to check your microscope is compatible.

If you are using transmitted light please go to the CONDENSER EXTENSION LENS section for condenser info.