LTS420E-P Specifications

  • Temperature Range < -195°C to 420°C (add LNP95 cooling pump to cool sample below ambient)

  • Sample area of 58x43mm

  • Quick-release gas valves for atmospheric control

  • Swing out lid for easy sample loading

  • Can be used with transmitted or reflected light

  • Mounts directly to microscope table or substage

  • Stage body size - 169 x 110 x 25.8 mm

  • 100 ohm platinum sensor

  • Temperature stability & accuracy to <0.01°C

  • Highly conductive metal for improved heat transfer

  • Maximum heating / cooling rate of 30°C/min

  • Response time of <1 second at 5°C/min at 50°C

  • Objective lens minimum working distance 8.2mm

  • Condenser lens minimum working distance 12.5mm